Application of thermography to estimate respiratory rate in the emergency room: The journal Temperature toolbox

Nenhuma Miniatura disponível
Citações na Scopus
0
Tipo de produção
article
Data de publicação
2023
Título da Revista
ISSN da Revista
Título do Volume
Editora
TAYLOR AND FRANCIS LTD.
Autores
ALDRED, A.
RIBEIRO, J. A. S.
BEZERRA, P. M. S.
DESUó, I. C.
GOMES, G.
Citação
TEMPERATURE, v.10, n.2, p.159-165, 2023
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
Among the vital signs collected during hospital triage, respiratory rate is an important parameter associated with physiological, pathophysiological, and emotional changes. In recent years, the importance of its verification in emergency centers due to the severe acute respiratory syndrome 2 (SARS2) pandemic has become very clear, although it is still one of the least evaluated and collected vital signs. In this context, infrared imaging has been shown to be a reliable estimator of respiratory rate, with the advantage of not requiring physical contact with patients. The objective of this study was to evaluate the potential of analyzing a sequence of thermal images as an estimator of respiratory rate in the clinical routine of an emergency room. We used an infrared thermal camera (T540, Flir Systems) to obtain the respiratory rate data of 136 patients, based on nostrils’ temperature fluctuation, during the peak of the COVID-19 pandemic in Brazil and compared it with the chest incursion count method, commonly employed in the emergency screening procedures. We found a good agreement between both methods, with Bland-Altman limits of agreement ranging from −4 to 4 min−1, no proportional bias (R2 = 0.021, p = 0.095), and a strong correlation between them (r = 0.95, p < 0.001). Our results suggest that infrared thermography has potential to be a good estimator of respiratory rate in the routine of an emergency room.
Palavras-chave
Bland-Altman, emergency room, pandemics, respiratory rate, SARS2, Thermography, triage
Referências
  1. Chester J.G., Rudolph J.L., Vital signs in older patients: age-related changes, J Am Med Dir Assoc, 12, 5, pp. 337-343, (2011)
  2. Rechtman E., Curtin P., Navarro E., Et al., Vital signs assessed in initial clinical encounters predict COVID-19 mortality in an NYC hospital system, Sci Rep, 10, (2020)
  3. Huang C., Wang Y., Li X., Et al., Clinical features of patients infected with 2019 novel coronavirus in Wuhan, China, Lancet, 395, pp. 497-506, (2020)
  4. Cretikos M.A., Bellomo R., Hillman K., Et al., Respiratory rate: the neglected vital sign, Med J Aust, 188, 11, pp. 657-659, (2008)
  5. Droitcour A.D., Seto T.B., Park B.-K., Et al., Non-contact respiratory rate measurement validation for hospitalized patients, Annu Int Conf IEEE Eng Med Biol Soc, 2009, pp. 4812-4815, (2009)
  6. Pereira C.B., Yu X., Czaplik M., Et al., Remote monitoring of breathing dynamics using infrared thermography, Biomed Opt Express, 6, 11, pp. 4378-4394, (2015)
  7. Ebell M.H., Predicting pneumonia in adults with respiratory illness, Am Fam Physician, 76, 4, pp. 560-562, (2007)
  8. Mochizuki K., Shintani R., Mori K., Et al., Importance of respiratory rate for the prediction of clinical deterioration after emergency department discharge: a single‐center, case–control study, Acute Med Surg, 4, 2, pp. 172-178, (2017)
  9. Molgaard R.R., Larsen P., Hakonsen S.J., Effectiveness of respiratory rates in determining clinical deterioration: a systematic review protocol, JBI Database Syst Rev Implement Rep, 14, 7, pp. 19-27, (2016)
  10. Bianchi W., Dugas A.F., Hsieh Y.-H., Et al., Revitalizing a vital sign: improving detection of tachypnea at primary triage, Ann Emerg Med, 61, 1, pp. 37-43, (2013)
  11. Zane R.D., Emergency department triage respiratory rate measurements are often inaccurate, NEJM J Watch, (2013)
  12. Acutely ill patients in hospital: recognition of and response to acute illness in adults in hospital. London: national Institute for Health and Clinical Excellence (UK), (2007)
  13. Philip K.E.J., Pack E., Cambiano V., Et al., The accuracy of respiratory rate assessment by doctors in a London teaching hospital: a cross-sectional study, J Clin Monit Comput, 29, pp. 455-460, (2015)
  14. Lovett P.B., Buchwald J.M., Sturmann K., Et al., The vexatious vital: neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triage, Ann Emerg Med, 45, 1, pp. 68-76, (2005)
  15. Edmonds Z.V., Mower W.R., Lovato L.M., Et al., The reliability of vital sign measurements, Ann Emerg Med, 39, 3, pp. 233-237, (2002)
  16. Abbas A.K., Heimann K., Jergus K., Et al., Neonatal non-contact respiratory monitoring based on real-time infrared thermography, Biomed Eng Online, 10, 1, (2011)
  17. Fei J., Pavlidis I., Thermistor at a distance: unobtrusive measurement of breathing, IEEE Trans Biomed Eng, 57, 4, pp. 988-998, (2010)
  18. Lewis G.F., Gatto R.G., Porges S.W., A novel method for extracting respiration rate and relative tidal volume from infrared thermography, Psychophysiology, 48, 7, pp. 877-887, (2011)
  19. Pan F., Yang L., Li Y., Et al., Factors associated with death outcome in patients with severe coronavirus disease-19 (COVID-19): a case-control study, Int J Med Sci, 17, 9, pp. 1281-1292, (2020)
  20. Bland J.M., Altman D.G., Measuring agreement in method comparison studies, Stat Methods Med Res, 8, 2, pp. 135-160, (1999)
  21. Bland J.M., Altman D.G., Statistical methods for assessing agreement between two methods of clinical measurement, Lancet, 327, 8476, pp. 307-310, (1986)
  22. Pereira C.B., Czaplik M., Blazek V., Et al., Monitoring of cardiorespiratory signals using thermal imaging: a Pilot Study on healthy human subjects, Sensors, 18, 5, (2018)
  23. Lorato I., Stuijk S., Meftah M., Multi-camera infrared thermography for infant respiration monitoring, Biomed Opt Express, 11, pp. 4848-4861, (2020)